EasyManuals Logo

Roche COBAS INTEGRA 400 plus User Manual

Roche COBAS INTEGRA 400 plus
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
66 Result blocks
Block 06: Control Result with Lot Information
Block 06 is transmitted to the host if all of the following
conditions are met:
After receiving a result request with result type
selector 11 or 19
If an accepted and not yet transmitted control result is
available
If the instrument is configured accordingly
Results always collated by test ID Result type selector 11 or 19 is used.
m = 1 (one test ID per result block)
n = 1 to 6
q Results of different control orders are not collated.
Line
code
Description Iteration
55 Test ID
1 to m
15 Access Date Time
16 Result Indexes
11 Reagent Lot
Information
02 Control ID
1 to n
12 Control Lot
Information
00 Result Data
y Block 06

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals