4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
66 Result blocks
Block 06: Control Result with Lot Information
Block 06 is transmitted to the host if all of the following
conditions are met:
• After receiving a result request with result type
selector 11 or 19
• If an accepted and not yet transmitted control result is
available
• If the instrument is configured accordingly
Results always collated by test ID Result type selector 11 or 19 is used.
• m = 1 (one test ID per result block)
• n = 1 to 6
q Results of different control orders are not collated.
Line
code
Description Iteration
55 Test ID
1 to m
15 Access Date Time
16 Result Indexes
11 Reagent Lot
Information
02 Control ID
1 to n
12 Control Lot
Information
00 Result Data
y Block 06