EasyManuals Logo

Roche COBAS INTEGRA 400 plus User Manual

Roche COBAS INTEGRA 400 plus
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #160 background imageLoading...
Page #160 background image
8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
160 List of line codes for extended HIF
Line Code 39: Cassette Status
Example
39_19_19/02/2016_13/05/2016<LF>
q Date opened and date expired will be rendered as
00/00/0000 if the information is not available.
Every status can be mapped to one of three colours (Red,
Yellow and Green). The following table shows the
possible returned status with corresponding meaning and
colour.
Description Format Range
Line Code 39
Status I2 Detailed information is
provided in the table below
Date opened D1 dd/mm/yyyy
Date expired D1 dd/mm/yyyy
y Line Code 39: Cassette Status
Status code Meaning Color
12 In use Green
19 New cassette Green
20 Ready Green
1 Expired Yellow
5 Not enough for worklist Yellow
8 Low Yellow
9 Near expiration Yellow
13 Calibration expired Yellow
14 Requires mixing Yellow
15 Reconstituting Yellow
33 Preparing Yellow
2 Empty Red
4 Not on board Red
6 Missing definition (unknown
fluid)
Red
y Status codes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals