6 Flags and error codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
108 Flag CALC (result error)
Flag CALC (result error)
Flag CALC provides information about the checks
performed during the calculation of the result.
q A flag with a higher priority (lower number)
supersedes all other flags.
Flag value Priority Name Meaning
0 No flag
1 1 HIGH ABS High absorbency
2 2 DARK ERR Dark error
3 3 OUTLIERS Outliers
4 4 FP UNSTAB FP unstable
5 5 POL ERR Polarization error
6 6 LOW SIGNAL Low signal intensity
7 7 NO FLUID No sample
8 8 NO ISECAL No calibrator bottle
9 9 SEG FLUID No end of sample segment detected
10 10 SEG ISECAL No end of standard segment detected
11 11 AIR FLUID Not enough sample / too many bubbles
12 12 AIR ISECAL Not enough standard / too many bubbles
13 13 ISE UNSTAB ISE measurement unstable
14 14 < REAG RNG Low reagent range exceeded
15 15 > REAG RNG High reagent range exceeded
16 16 EP UNSTAB Endpoint unstable
17 17 HIGH ACT High activity (please refer to following note)
18 18 LOW ACT Low activity (please refer to following note)
19 19 CURVE DIR Curve direction
20 20 NON LINEAR Non-linear kinetic
21 21 < KIN READ Insufficient kinetic readings
22 22 HIGH BKG High background noise (please refer to following note)
23 23 > REPL DEV Replicate deviation check failed
24 24 CALC ERROR Calculation error (please refer to following note)
25 25 NON MONO Curve not monotone
26 26 OUT OF RNG Curve / slope out of range
27 27 > STD DEV Standard deviation exceeded
28 28 SOL 1 F DEV Solution 1 factor deviation
29 29 AG EXCESS Antigen excess
30 30 < TEST RNG Lower test range exceeded (please refer to following note)
31 31 > TEST RNG Upper test range exceeded (please refer to following note)
255 33 TEST BLOCKED Used with collated results if a test is blocked (see note)
y List of flag CALC

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals