6 Flags and error codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Flags and error codes 107
Sub-flag R = (Flag X and 12)/4
Sub-flag ^ = Flag X and 3
Value Name Meaning
0 no flag Result is error-free.
1 r The result is erroneous or corrected, or
the result of the replicate is erroneous.
The result of the first measurement and
the replicate were accepted.
2 R The result is erroneous or corrected, or
the result of the replicate is erroneous.
The result of the first measurement and
the replicate were not accepted.
3 d The result was deleted or all the results
(with calibration measurements or
replicates) have been deleted.
y Sub Flag R = (Flag X and 12)/4
Value Name Meaning
0 no flag Result was generated without a
replicate, sample concentration or
sample dilution.
1 -> Result was generated with a replicate.
2 v Result was generated after dilution.
3 ^ Result was generated after
concentration.
y Sub Flag ^ = Flag X and 3