6 Flags and error codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
106 Flag X
Flag X
Flag X provides information about the conditions under
which a result was generated.
Flag X contains four subgroups.
•C
•Q
•R
•^
Flag X is transmitted as 3 ASCII characters which
represent a 3-digit integer value (value range 0 to 255).
This value represents an 8-bit value. The four subgroups
are coded within the four two-bit groups of a byte. The
single flags are binary coded within each two-bit group.
Sub-flag C = (Flag X and 192)/64
Sub-flag Q = (Flag X and 48)/16
Flag X
Decimal 0 ... 255
Binary 1 1 1 1 1 1 1 1
Sub-flag C Q R ^
y Flag notation
Value Name Meaning
0 no flag Result was calculated with an error-free
calibration.
1 c The result was calculated with an
erroneous and/or corrected calibration.
The calibration was already accepted.
2 C The result was calculated with an
erroneous and/or corrected calibration.
The calibration was not accepted.
3 p The result was calculated with a
postponed calibration (I400 V2.0 only).
y Sub Flag C = (Flag X and 192)/64
Value Name Meaning
0 no flag The result was secured with an error-
free control.
1 q The result was secured with an
erroneous and/or corrected control. The
control was accepted.
2 Q The result was secured with an
erroneous and/or corrected control. The
control was not accepted.
3 Not used
y Sub Flag Q = (Flag X and 48)/16