4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
64 Result blocks
Block 03: Control Result
Block 02 is transmitted to the host if all of the following
conditions are met:
• After receiving a result request
• If an accepted and not yet transmitted control result is
available
• If the instrument is configured accordingly
Results collated by test ID Result type selector 01 or 09 is used.
• m = 1 (one Test ID per result block)
• n = 1 to 6
q Results of different control orders are not collated.
Results collated by control ID Result type selector 06 is used.
• m = 1 to # of available tests (m is the number of tests
with not yet transmitted QC results)
• n = 1 (one control ID per result block)
Block 04: Patient Results
Block 04 is transmitted to the host if all of the following
conditions are met:
• After receiving a result request block 09
• If an accepted and not yet transmitted patient result is
available
Line
code
Description Iteration
55 Test ID
1 to m
01 Result Time
02 Control ID
1 to n
00 Result Data
y Block 03
Line
code
Description Iteration
(1)(2)
(1) n = 1. If result type selector 01 or 07 is used (single result)
(2) n = 1 to 50. If result type selector 03 or 04 is used (results collated
by order number)
53 Order ID
56 Sample Name 0 to 1
55 Test ID
1 to n
00 Result Data
y Block 04