8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface 163
Line Code 49: Rack Position
Example
49_H_2-10<LF>
The first character is either the SLOT_ID letter (A-Z) or a
whitespace if the slot is not on board. It is then followed
by [rack number]-[rack position].
Line code 55: Test ID
Example
55_••1_•••2.00_•••3.50<LF>
q Profile and ratio handling: Use ratio number to order
ratio.
Ratio and individual tests results will be delivered.
The available test no., profile IDs and Ratio IDs can be
found in Configuration > Test Definitions > ID.
Samples in the instrument can only be diluted. The
maximum dilution factor is 100.
If the dilution factor is omitted, the 1:1 factor will be used.
Description Format Range
Line Code 49
Rack position S10 SlotID_RackNumber-RackPosition
y Line Code 49: Rack Position
Description Format Range
Line code 55
Test no. I3 1 to 999
Optional If dilution factor is extra information for block
10
Dilution factor F7 1 to 100
Automatic Rerun
Dilution factor
F7 1 to 100
y Line code 55: Test ID

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals