8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface 157
Line Code 29: Cassette status
Example
29_07-6831-6••••••_•1_G•7-1<LF>
Line Code 30: Operation selection
Example
30_1<LF>
Line Code 31: Calibrator status
Example
31_07-6830-8••••••_CFAS••••_•6_•1_J•13-6••••<LF>
Description Format Range
Line Code 29
Cassette ID S15 Number ID of the cassette
Cassette Status I2 For detailed information, refer to table displayed in
Line 39.
Optional If the rack position is available
Rack Position S10 SlotID_RackNumber-RackPosition
y Line Code 29: Cassette status
Description Format Range Range description
Line Code 30
Operation setting I1 1 Start the
instrument
2 Stop the
instrument
y Line Code 30: Operation selection
Description Format Range Range description
Line Code 31
Calibrator ID S15 Number ID of the calibrator
Calibrator Short Name S8 Short name of the calibrator
Std I2 Standard number of the calibrator lot
Calibrator status I2 0 OK
1 Expired
2 Empty
3 Blocked
4 Not on board
22 Calibration Pending
Optional If the rack position is available
Rack Position S10 SlotID_RackNumber-RackPosition
y Line Code 31: Calibrator status