5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes 91
Line code 16: Result Indexes
Example calibration
16_1141901502_••••••••••_•••••••••0<LF>
Example control
16_1141901502_1141901523_•••••••••0<LF>
Example sample
16_1141901502_1141901523_1141901678<LF>
Example sample without control
16_1141901502_•••••••••0_1141901678<LF>
q A calibration result comprises only the calibration
index with which the calibration itself will be identified.
q A control result is comprised of the calibration index
(reference to the calibration with which the control was
calculated) and a control group index, identifying the
control itself.
q A sample result is comprised of:
o A calibration index (reference to the calibration with
which the sample result was calculated)
o If there is an associated control, then there is a control
group index (reference to the control).
o A sample index, with which the result itself will be
identified
q A result will have a maximum of three indexes. If one
index does not exist, it will be replaced with a right-
justified zero. See the examples above.
Description Format Range
Line code 16
Calibration index S10
Control group index S10
Sample result index S10
y Line code 16: Result Indexes