5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes 91
Line code 16: Result Indexes
Example calibration
16_1141901502_••••••••••_•••••••••0<LF>
Example control
16_1141901502_1141901523_•••••••••0<LF>
Example sample
16_1141901502_1141901523_1141901678<LF>
Example sample without control
16_1141901502_•••••••••0_1141901678<LF>
q A calibration result comprises only the calibration
index with which the calibration itself will be identified.
q A control result is comprised of the calibration index
(reference to the calibration with which the control was
calculated) and a control group index, identifying the
control itself.
q A sample result is comprised of:
o A calibration index (reference to the calibration with
which the sample result was calculated)
o If there is an associated control, then there is a control
group index (reference to the control).
o A sample index, with which the result itself will be
identified
q A result will have a maximum of three indexes. If one
index does not exist, it will be replaced with a right-
justified zero. See the examples above.
Description Format Range
Line code 16
Calibration index S10
Control group index S10
Sample result index S10
y Line code 16: Result Indexes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals