5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
86 List of line codes
Line code 03: Standard Rates
Example
03_C_-1.929185E-02_-1.513617E-02<LF>
Line code 04: Calibration Curve
Example
04_•1.000000E+00_-1.513617E-02_-5.716211E-04<LF>
The meaning of parameters 1 to 5 depends on the
calibration mode used:
• For qualitative calibration, parameter 1 corresponds to
the cut-off value.
• For quantitative calibration, parameters 1 to 5
correspond to the curve parameters of the model
used.
• For ISE calibration, the measured electrode defines
the parameter. If the Li electrode has been measured,
parameter 1 contains the Li slope, parameter 2 the Na
slope and parameter 3 the K slope.
• For all other electrodes, parameter 1 contains the
associated slope value.
Description Format Range Range description
Compulsory Line code 03
Standard type S1 C Calibration (mean rate)
O Correction (mean rate)
A Cal/Cor antigen excess rate
F13 +/-9.999999E+99 to +/-1.0E-99
Optional Standard rate F13
Standard rate F13
Standard rate F13
Standard rate F13
Standard rate F13
y Line code 03: Standard Rates
Description Format Range
Compulsory Line code 04
Correction factor F13 +/-9.999999E+99 to +/-1.0E-99
Correction offset F13 +/-9.999999E+99 to +/-1.0E-99
Correction parameter 1 F13 +/-9.999999E+99 to +/-1.0E-99
Optional Correction parameter 2 F13
Correction parameter 3 F13
Correction parameter 4 F13
Correction parameter 5 F13
y Line code 04: Calibration Curve