5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes 97
Results:
o If in Configuration Processing > Result Output
> Sample Type, the sample type is disabled, no
sample type information is transmitted to the host.
Ratio orders:
o If the component results of a ratio are measured from
different sample types, the Order Entry Sample Type
must be “???”.
u Related topics
Configuring barcode setup (192)
Line code 11: Reagent Lot Information (89)
Line code 54: Order Information
Example
54_••4_•4_S_These•are•21•chars•••<LF>
Description Format Range Range description
Compulsory Line code 54
Sample position rack number I3 0 Defined by barcode
1 to 999
Sample position tube number I2 0 Defined by barcode
1 to 15
Order priority S1 A Active
SSTAT
Optional Depending on database set-up parameters Leadertext 1 and 2
Sample leadertext 1 S21
Sample leadertext 2 S21
y Line code 54: Order Information

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals