7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes for extended host interface 133
Message Log blocks
In this section
Block 75: Message Log Request (133)
Block 76: Message Log Response (133)
Block 77: Failed Message Log Request (133)
Block 75: Message Log Request
It is possible to retrieve all the messages received by the
control unit after the specified date and time.
Block 76: Message Log Response
This request provides all the messages in the message
log received after date sent with block 75.
Block 77: Failed Message Log Request
Block 77 contains Line 96 and provides the eventual error
encountered following request with Block 75.
Line code Description
24 Date and Time
y Block 75
Line code Description Iteration
(1)(2)
(1) m is the total number of messages received after the specified date
and time.
(2) n is the total length of the message string divided by 100 (rounded
to the next integer).
24 Date and time
1 to m82 Message Icon
83 Message String 1 to n
y Block 76
Line code Description
96 Error Message
y Block 77