7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes for extended host interface 133
Message Log blocks
In this section
Block 75: Message Log Request (133)
Block 76: Message Log Response (133)
Block 77: Failed Message Log Request (133)
Block 75: Message Log Request
It is possible to retrieve all the messages received by the
control unit after the specified date and time.
Block 76: Message Log Response
This request provides all the messages in the message
log received after date sent with block 75.
Block 77: Failed Message Log Request
Block 77 contains Line 96 and provides the eventual error
encountered following request with Block 75.
Line code Description
24 Date and Time
y Block 75
Line code Description Iteration
(1)(2)
(1) m is the total number of messages received after the specified date
and time.
(2) n is the total length of the message string divided by 100 (rounded
to the next integer).
24 Date and time
1 to m82 Message Icon
83 Message String 1 to n
y Block 76
Line code Description
96 Error Message
y Block 77

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals