6 Flags and error codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
114 List of line code 99 error codes
List of line code 99 error codes
Error code Meaning Error condition
01 Block check sum error Transmission error
02 Parity error Transmission error
03 Line too long (>128 characters) Erroneous line configuration, or possibly transmission error if
block check off
04 Block too long Erroneous block configuration, or possibly transmission error if
block check off
05 Invalid message configuration Erroneous message configuration, or possibly transmission
error if block check off
06 Invalid block code Invalid block code, or possibly transmission error if block
check off
07 Invalid line code Invalid line code, or wrong mode used (patient/sample), or
possibly transmission error if block check off
08 Line format error Erroneous line configuration, or possibly transmission error if
block check off
y Line code 99 error codes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals