5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes 85
or
"<TEST RNG" flag, HIF returns the value and the lower
range limit:
00_•3.694475E+01_U/l•••_••
0_••0_•30_••0_•1.500000E+02_•5.000000E+00<LF>
q Result will contain the concentration value, even if
the flag is ">TEST RNG" or "<TEST RNG".
Range limit Contains the upper limit in the case that we get a
">TEST RNG" flag.
Contains the lower limit in the case that we get a
"<TEST RNG" flag.
Contains 0.000000E+00 in any other case (flag or not).
Line code 01: Result Time
Example
01_22:07:59<LF>
Line code 02: Control ID
Example
02_ROC-N<LF>
Description Format Range
Line code 01
Time T1 hh:mm:ss
y Line code 01: Result Time
Description Format Range
Line code 02
QC-name S5
y Line code 02: Control ID