8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
154 List of line codes for extended HIF
Line Code 25: Test Order Status
Example
25_131_6_24<LF>
Description Format Range
Line Code 25
Test ID I3 1 to 999
Test Status I1 Refer to following
table for details
Test Blocked Status I2 Refer to following
table for details
y Line Code 25: Test Order Status
Status flag code Description
0 New
1 To be rerunned
2 Calculated
3 Running
4 Finished
5 Failed
6 Blocked
y Status flag codes
Blocking status
flag code
Description
0 Okay
1 No fluid in sample cup
2 Reagent not on board
3 Reagent is empty
4 No cassette definition found
5 Diluent not on board
6 Diluent is empty
7 No diluent definition found
8 No calibrator on board
9 Calibrator empty
10 No valid calibration data
11 No calibration defintion available
12 No lot values for calibration available
13 Last calibration is not accepted
14 Control not on board
15 Control empty
16 Control is not valid
17 No control definition available
18 No assigned values for control available
19 Last control not accepted
20 Repeat action not possible
21 Test disabled
y Blocking status flag codes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals