6 Flags and error codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
112 List of line code 96 error codes
List of line code 96 error codes
Error code Host request Error condition
00 Any Manipulation OK (i.e. no error, manipulation accepted)
11 Patient Modify/Delete Patient ID not found
Order Entry Patient ID not found
12 Patient Entry Patient ID already exists
13 Pat.Entry/Mod./Delete Patient already locked (e.g. by work area etc.) or database
time-out
Order Entry Database time-out
14 Patient Deletion Patient still has open orders
21 Order Entry (Add Test) Sample ID (Order No./Sample Type/Order Date) not found
Order Deletion Sample ID (Order No./Sample Type/Order Date) not found
Change Work List Priority Attempt to change orders to passive priority
Change Work List Priority No appropriate orders found to change priority
Remove Tests from Order No appropriate tests found in order; No tests removed
22 Order Entry Sample ID already exists; rack/sample position already in use
Order Entry (Add Test) Sample order full (50 test orders already assigned)
Special Order Entry Calibration/Control request already set
23 Any Order Manipulation Order already locked (e.g. by work area, etc.), or
Database time-out
24 Any Order Manipulation Test not defined
25 Order Deletion Order contains not-yet accepted results
33 Result Request Result data already locked
Or: any database error such as:
o Sample type not found in sample type list
o Result has just been deleted
o Database time-out
o etc.
40 Operation Setting No communication with real time software
System Status Request No communication with real time software
Service Action Execution No communication with real time software
41 Operation Setting The current instrument status does not allow the required
change
Service Action Execution The current instrument status does not allow the requested
action
52 Cassette Status No cassette on board
55 Sample Order Status No pending tests
56 Sample Order Status Selected Order is not present
58 Test Status Request No tests loaded
59 Test Status Request The specified test is not loaded
61 MCS Pend. Sample Tubes No tubes found
63 MCS Pend. Sample Tubes Sample type not found in sample type list
MCS Calibr./Contr. status No test defined
Any MCS Request Database time-out
65 Calibration Status No test is loaded
66 Service Action Status No Service Action definition is available
y Line code 96 error codes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals