8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
162 List of line codes for extended HIF
Line Code 46: Currently Time or Tests elapsed
Example
46_••••••1230_•••••••••3<LF>
Line Code 47: Service Due
Example
47_••••••2000_•••••••••5<LF>
Line Code 48: Elapsed percentage
Example
48_••••••••40_•••••••••32<LF>
Description Format Range
Line Code 46
Counts of Time or Tests Elapsed I10
Days since last service action I10
y Line Code 46: Currently Time or Tests elapsed
Description Format Range
Line Code 47
Hours or Tests that must pass before
service action is performed
I10
Number of days that must pass before
service action is performed
I10
y Line Code 47: Service Due
Description Format Range
Line Code 48
Percentage of Hours or Tests occurred
for service action to be performed
I10
Percentage of days occurred for service
action to be performed
I10
y Line Code 48: Elapsed percentage