5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes 93
Line code 42: Tube Information
Example
42_•21_25_1_AB(012:3456)•••_URI<LF>
Line code 43: Test State
Example
43_•32_1<LF>
Description Format Range Range description
Compulsory Line code 42
Sample position rack number I3 0 Defined by barcode
1 to 999
Sample position tube number I2 0 Defined by barcode
1 to 15
Tube type I1 0 Fixed position tube
1 Barcode tube
Order number S15 Barcode characters
Optional If Sample Types Used is enabled in the database set-up, then sample type is included in line 42.
Sample type S3 Laboratory defined string.
Barcode characters
y Line code 42: Tube Information
Description Format Range Range description
Line code 43
Test ID I3 1 to 999
Test state I1 0 Test unavailable
(blocked)
1 Test available
y Line code 43: Test State