5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
94 List of line codes
Line code 44: Cal/CS State
Example
44_C_A_10:32:45_30/11/1993<LF>
Line code 50: Patient ID
Example
50_123.12.123.••••<LF>
WARNING
!
Incorrect result assignment due to use of same
patient ID (if in patient mode)
Patient ID has to be unique.
r Do not use the same Patient ID for more than one
patient, otherwise results could be assigned to the
wrong patient.
u Related topics
• Patient mode (30)
Description Format Range Range description
Line code 44
CAL/CS state type S1 C C = Calibration state
O Correction state
Q Control state
CAL/CS state S1 A Accepted
C Calculated
P Pending
R Reactivated
F Failed (calibrations and corrections only)
Time of last Cal/CS issue T1
Date of last Cal/CS issue D1 01/01/1800 to 31/12/2100
y Line code 44: Cal/CS State
Description Format Range
Line code 50
Patient ID S15
y Line code 50: Patient ID