EasyManuals Logo

Roche COBAS INTEGRA 400 plus User Manual

Roche COBAS INTEGRA 400 plus
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes for extended host interface 127
Service Action Blocks
In this section
Block 25: Service Action Execution (127)
Block 26: Service Action response (127)
Block 27: Service Action Result Request (127)
Block 28: Service Action Result Response (128)
Block 29: Failed Service Action Result Response (128)
Block 25: Service Action Execution
Automatic Service Action can be performed from host. In
order to perform the automatic service actions the system
needs to be in Standby mode. The service action is
specified by Line 32.
Block 26: Service Action response
Each request with Blocks 25 is answered by Block 26 and
Line 96.
The response message depends on the success of the
request.
Block 27: Service Action Result Request
It is possible to retrieve all service action execution results
after the specified date and time.
Line code Description
32 Service Action
y Block 25
Line code Description
96 Error Message
y Block 26
Line code Description
24 Date and time
y Block 27

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals