8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
144 List of line codes for extended HIF
Line Code 17: Test Execution
Example
17_23/04/2006_22:07:59_1:1•••_••0<LF>
q oExecution Date and Time is the point when the
measurements are executed on the Instrument (Calculation
Date Time)
o If the dilution flag is shorter than six characters, the
string will be filled with spaces on the right side.
Example, 1:1 will be 1:1•••
(•) represents a filler space
o If the dilution flag is shorter than three characters, the
string will be filled with spaces on the left side.
Example, 0 will be••0
(•) represents a filler space
Description Format Range Range description
Line Code 17
Date D1 dd/mm/yyyy
Time T1 hh:mm:ss
Dilution Flag S6 1:X with X going from
1 to 100
Sample Flag I3 Refer to following
table
y Line Code 17: Test Execution
Flag code Description
13 Result calculated using an erroneous or corrected calibration which has been accepted. (F1)
69 Result calculated using a postponed calibration which has been accepted. Note: Postponing a
calibration may lead to erroneous results. Action: Review results - Repeat calibration if necessary.
32 Result is associated with an erroneous or corrected control. The control has been accepted. (F1).
34 Result (or a replicate) is erroneous or corrected. Result has been accepted. Recommended action:
Check results reaction curve to establish the cause of the deviation; repeat sample measurement (F1)
10 (Serum proteins only) The sample contains an excess of antigen. %1% = [value] - is Test Range High
(F3)
12 Insufficient calibrator or too many air bubbles. The calibrator segment is too short or air bubbles are
present, or a tube is leaking or blocked. (F3)
11 Insufficient amount of sample/QC/Solution or too many air bubbles in electrode path. The sample
segment is too short or air bubbles are present, mixing is too vigorous (in the case of the indirect/urine
mode) or a tube is leaking or blocked. (F3)
255 Any test flag.
15 Calculation error from pre-calculation (Abs / Fp raw data)
65 Result obtained using a concentration factor. (F1)
17 The dark measurement was higher than the test measurement. (F3)
62 Result differs from last result (up to n days before) by greater than defined %. %1% = [def. value] in
percent (F4)
61 Result differs from last result (up to n days before) by greater than defined value. %1% = [def. value]
%2% = [unit] (F4)
64 Result obtained using a dilution factor. (F1)
y Flag codes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals