4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes 61
About block codes
The following is the table of contents for the block code
chapter. It also shows the request/response behavior for
host-instrument communication.
Block composition notation
Host (request) Instrument (response)
Block code Request Block code Request
00 Idle Block / Synch Block 00 Idle Block
09
.
Result Request 02 Calibration Result
03 Control Result
04 Patient Result
05 Calibration Result with Lot Information
06 Control Result with Lot Information
07 Patient Result with Lot Information
08 Result Request Response
00 Idle Block, if no result available
10 Order Entry 19 Order Manipulation Response
11 Order Deletion
40 Patient Entry 49 Patient Manipulation Response
41 Patient Deletion
42 Patient Modification
60 Multi Configuration Service Request 61 Slot Configuration
62 Sample Tube Inform.
63 Cal-CS State
69 Service Request Response
90 System Status Request 91 System Status Response
92 Protocol Version Request 93 Protocol Version Data
.- any Host Request 99 Control Message
y List of block codes
Line
code
Description Iteration
55 Test ID
1 to m
(1)
(1) Number of iterations in all lines
05 Access Date Time
16 CALidx
11 Reagent Lot
Information
02 Control ID
1 to n
(2)
(2) Number of iterations in lines 02 and 12
12 Control Lot
Information
y Notation of block codes