8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface 161
Line code 40: Service Selection
Example
40_1<LF>
Line Code 45: Service Action Definition
Example
45_103_c<LF>
Description Format Range Range description
Line code 40
Service request I1 0 Slot configuration
1 Sample tubes without
orders
3 Reserved
4 List of all sample tubes
5 Status of all automatic
service actions
y Line code 40: Service Selection
Description Format Range Range description
Line Code 45
Service Action ID I3
Counts unit S1 h counts are given in
hours
c counts are given in
test counts
y Line Code 45: Service Action Definition
ID Description
0 Deproteinize probes
101 Prime fluid system
103 Backup database
301 Electrode service
304 Activate electrodes
310 Prime ISE calibrators
312 Clean ISE tower automatically
314 Condition ISE tubing
315 Initialize ISE module
y IDs

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals