8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
170 List of line codes for extended HIF
Line Code 97: System status
Example
97_•5_0_0_0_1_•853_0_•••37.1_•••12.6_•••24.2<LF>
Description Format Range Range description
Line Code 97
System status I2 0 Sleeping
1 Initializing
2 Stopped
3 Service
4 Cover Open
5 Standby
6 Running
7 Fatal Error
8 Power Down
9 Auto Service
10 Power On
Cleaner status I1 0 Ok
1 Warning
2 Error
Cuvette status I1 0 Ok
1 Warning
2 Error
Water supply
status
I1 0 Ok
1 Warning
2 Error
Cuvette waste
status
I1 0 Ok
1 Warning
2 Error
Cuvette waste
count
I4 0..1000
Waste drain
status
I1 0 Ok
1 Warning
2 Error
Analyzer
temperature
F7 Temperature in degrees
Cassettes
temperature
F7 Temperature in degrees
Cleaner
temperature
F7 Temperature in degrees
y Line Code 97: System status