2 About host interface features
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
44 Result requests
Result requests
In this section
About result requests (44)
Example of result query (46)
Example of unspecific result request (48)
Example of result request with ratio (collated result
request) (49)
About result requests
Due to the request/response principle, results are
transmitted to the host only upon the receipt of a Result
Request (BC09).
COBAS INTEGRA 400 plus analyzer and
Cedex Bio HT Analyzer allow requesting single results or
collated results. The BC09/LC10 thus contains a result
type selector.
q We strongly advise against the use of collated
results, depending on the database size or fragmentation,
we cannot guaranty the instrument performance.
No specific result can be requested.
Only accepted results will be transmitted. (Auto accept
can be configured in Configuration > Processing
> Result Handling Table).
Usually each result will be transmitted only once. To re-
transmitted a result:
On the instrument user interface, change to the result
workbench.
Select the required results with the mouse.
Select Retransmit in the Process menu.
We recommend using the following procedure and timing
for the result request loop:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals