4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes 67
Block 07: Patient Results with Lot Information
Block 07 is transmitted to the host if all of the following
conditions are met:
• After receiving a result request with result type
selector 11 or 17
• If an accepted and not yet transmitted patient result is
available
Block 08: Result Request Response
A possible error of a result request will be answered by
block 08.
q The message contains the associated error code.
Line
code
Description Iteration
53 Order ID
56 Sample Name 0 to 1
15 Access Date Time
16 Result Indexes
55 Test ID
11 Reagent Lot Information
00 Result Data
y Block 07
Line
code
Description
96 Error Message
y Block 08