8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface 159
Example
36_39901•••••_12398201••_07-6494-9•<LF>
Line Code 37: Tests left
Example
37_•••0_•••0_•85.5<LF>
q If the cassette is a cleaner the Tests left and Total
tests fields will be set to 0.
Line Code 38: Diluent status
Example
38_07-5635-0••••••_•1_O•140-8•••<LF>
Description Format Range
Line Code 37
Tests left I4 The number of tests left for the specific cassette
Total tests I4 The total number of tests originally available for the
specific cassette
Optional If the cassette is a cleaner
Percentage of filling F5 The filling percentage when the cassette is a
cleaner.
y Line Code 37: Tests left
Description Format Range Range description
Line Code 38
Diluent ID S15 Number ID of the diluent
Diluent status I2 0 OK
1 Expired
2 Empty
4 Not on board
6 Missing information
8Low
9 Near expiry
12 In use
19 New
20 Ready
33 Preparing
Optional If the rack position is available
Rack Position S10 SlotID_RackNumber-RackPosition
y Line Code 38: Diluent status

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals