Line codes for extended host interface 143
Line Code 08: FP Raw Data
Control
08_114190150_114190152_•••••••••0<LF>
q Notes regarding the results replicates:
The 4 fields will be replicated r times in the same Line.
r = Total number of replicates
Line Code 09: ISE Raw Data
Example
09_••78.20<LF>
Line Code 14: Validation Status
Example
14_1<LF>
Description Format Range
Line Code 08
Blank parallel I9 1 to r
Blank vertical I9
Test parallel I9
Test vertical I9
y Line Code 08: FP Raw Data
Description Format Range$
Line Code 09
ISE Mean Value F7
y Line Code 09: ISE Raw Data
Description Format Range Range description
Line Code 14
Validation Status I1 0 Accepted
1 Disabled
2 Rerun
3 Not handled
y Line Code 14: Validation Status