5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
96 List of line codes
Line code 53: Order ID
Example
53_abcdef987654321_00/00/0000_URI<LF>
q The order number and the sample type may not
contain <space>s in front of or within the identifier.
Nevertheless, if the fields are not used in their full length,
digits left over have to be filled with <space>s.
Predefined sample types There are three generally predefined sample types:
WARNING
!
r Using a sample specific code instead of “???” (for
example HEM for hemolysate) will override any
instrument and test default sample settings and force
all tests from the order to be executed on such a
sample type. This has the same effect as selecting a
specific sample type on the order entry screen of the
instrument, instead of automatic, which is the default
in the drop down list.
q Orders:
o If the sample type is not transmitted, the default
sample type (defined by test or rack definition) is
used.
Description Format Range
Compulsory Line code 53
Order number S15 Barcode characters
Order date D1 01/01/1800 to 31/12/2100
00/00/0000 (= today)
Optional If sample type is used
Sample type S3 Laboratory defined string barcode characters
y Line code 53: Order ID
??? Sample type determined by instrument’s test
definitions. Used with order entries and deletions if
the test-specific sample type should be selected
automatically.
*** These wild cards are used with order deletions to
select all sample types.
RTO The instrument sets sample type to RTO for ratio
results.
RTO is not used when using collated results, but
still used with unspecific result requests (10 01)