4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
76 Multi-configuration service blocks
Multi-configuration service blocks
In this section
Block 60: Multi-Configuration Service (MCS) Request
(host query) (76)
Block 61: Slot Configuration (76)
Block 62: Sample Tube Information (77)
Block 69: Service Request Response (78)
Block 60: Multi-Configuration Service (MCS) Request
(host query)
Block 60 has one line.
Block 60 provides requests for the following functions:
Block 61: Slot Configuration
The reply to an MCS request with service = 0 is block 61
with the current slot configuration.
The rack numbers of the racks in the slots are sent.
Line
code
Description
40 Service Selection
y Block 60
0 Slot configuration f response: block 61
1 Onboard sample tubes without orders (host query mode)
f response: block 62
This request is used when working in host query mode
with a barcode on each sample tube.
2 Cal/CS status per test f response: block 63
3 Reserved
4 List of all known sample tubes > response: block 62
Line
code
Description
41 Slot State
y Block 61

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals