8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
146 List of line codes for extended HIF
Line Code 18: Cassette Information
Example
18_62159401••_68354•••••_12/05/2016_06/11/2017<LF>
q oIf no cassette lot number or serial number are found or
they were not defined, "??????????" will be used and
transmitted (without quotes).
o If no on board expiration date or cassette lot
expiration date are given for the cassette,
"00/00/0000" will be used and transmitted (without
quotes).
o If the lot number or serial number are shorter than ten
characters, the string will be filled with spaces on the
right side.
Example, 62159401 will be 62159401••
(•) represents a filler space
66 turbitidy sample present The sample has a cloudy color. (F2)
14 Result calculated using an erroneous or corrected calibration which has not been accepted. (F1)
70 Result calculated using a postponed calibration which has not been accepted. Note: Postponing a
calibration may lead to erroneous results Action: Check carefully the results - Repeat calibration if
necessary.
33 Result is associated with an erroneous or corrected control. The control has not been accepted.
Recommended action: repeat measurement with a fresh control; repeat sample measurement if
necessary. (F1).
35 Result (or a replicate) is erroneous or corrected. Result has not been accepted. Recommended action:
Check results reaction curve to establish the cause of the deviation; repeat sample measurement. (F1)
Flag code Description
y Flag codes
Description Format Range
Line Code 18
Cassette lot number S10
Cassette serial number S10
Cassette on board expiration
date
D1 dd/mm/yyyy
Cassette lot expiration date D1 dd/mm/yyyy
y Line Code 18: Cassette Information