5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
92 List of line codes
Line code 40: Service Selection
Example
40_1<LF>
Line code 41: Slot State
Example
41_••3_350_••0_••0_••0_784<LF>
Description Format Range Range description
Line code 40
Service request I1 0 Slot configuration
1 Sample tubes without
orders
3 Reserved
4 List of all sample tubes
y Line code 40: Service Selection
Description Format Range
Line code 41
Rack number in slot 1 I3 1 to 999
Rack number in slot 2 I3 1 to 999
Rack number in slot 3 I3 1 to 999
Rack number in slot 4 I3 1 to 999
Rack number in slot 5 I3 1 to 999
Rack number in slot 6 I3 1 to 999
y Line code 41: Slot State