EasyManuals Logo

Roche COBAS INTEGRA 400 plus User Manual

Roche COBAS INTEGRA 400 plus
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
84 List of line codes
List of line codes
Line code 00: Result Data
Examples Normal result without any flag:
00_•8.694475E+01_U/l•••_••
0_••0_••0_••0_•0.000000E+00_•0.000000E+00<LF>
or
">TEST RNG" flag, HIF returns the value and the higher
range limit:
00_10.694475E+01_U/l•••_••
0_••0_•31_••0_•1.500000E+02_•10.00000E+00<LF>
Line code In Line code In
00 Result Data 43 Test State
01 Result Time 44 Cal/CS State
02 Control ID 50 Patient ID
03 Standard Rates 51 Patient Information
04 Calibration Curve 52 Special Order Selection
07 ABS Sample Check 53 Order ID
10 Result Type Selection 54 Order Information
11 Reagent lot information 55 Test ID
12 Control Lot Information 56 Sample name
13 Calibration Lot Information 57 Priority Modification
15 Access Date Time 96 Error Code
16 Result Indexes 97 System Status
40 Service Selection 98 Protocol Version
41 Slot State 99 General Error Code
42 Tube Information
y List of line codes
Description Format Range
Line code 00
Result F13 +/-9.999999E+99.. +/-1.0E-99
Unit S6
Flag X I3 0..255
Flag S I3 0 (not used, currently set to 0)
Flag CALC I3 0..255
Flag QC I3 0..255
Range value to flag F13 +/-9.999999E+99.. +/-1.0E-99
Range limit F13 +/-9.999999E+99.. +/-1.0E-99
y Line code 00: Result Data

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals