8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface 153
Line Code 22: Order ID
Example
22_abcdef987654321<LF>
Line Code 23: Sample Order Status
Example
23_3<LF>
Line Code 24: Message Date Time
Example
24_23/04/2006_22:07:59<LF>
q The messages are not cleared via host interface so,
after a certain time, they should be deleted at the control
unit.
Description Format Range
Line Code 22
Order number S15 Sample Order ID
Optional If sample name information is present
Sample Name S50
y Line Code 22: Order ID
Description Format Range Range description
Line Code 23
Status I2 3 blocked
4 not on board
7 no orders
10 calculated
19 new
30 running
y Line Code 23: Sample Order Status
Description Format Range
Line Code 24
Date D1 dd/mm/yyyy
Time T1 hh:mm:ss
y Line Code 24: Message Date Time

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals