Roche Diagnostics
xii Operators Manual · Version 1.0
URISYS 2400
Radio interference
URISYS 2400 is a class A device. In residential areas it may cause radio
interference. The user must take precautions as required.
Installation
Follow the specified installation conditions carefully. Otherwise, inaccurate results,
or damage to the analyzer may occur.
Waste
Handle liquid waste and used test strips properly, according to legislation on water
pollution, and on the treatment of drainage and waste matter.
Electromagnetic waves
Devices that emit electromagnetic waves may affect measured data, or cause the
analyzer to malfunction. Do not operate the following devices in the same room
where the analyzer is installed: mobile phone, transceiver, cordless phone, other
electrical devices that generate electromagnetic waves.
o
Operating and maintenance
Carefully follow the procedures specified in the Operators Manual for the
operation and maintenance of the analyzer. Leave maintenance of other areas to
trained Roche Technical Support personnel.
o
Calibration
Calibrate the analyzer every four weeks to ensure the analyzer operates
optimally. Failure to do so may lead to inaccurate results.
o
Test strips
Handle and store test strips according to the instructions provided by Roche
Diagnostics. Refer to the instructions in the package insert of the URISYS 2400
cassette.
o
Particular care must be taken when performing the following procedures:
Preparing and processing samples (routine, STAT and control)
Using the washing solution
Draining the liquid waste tank
Cleaning the TS transfer base
Disposing of used test strips
Cleaning the waste box