URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
34 Reflectance Setting?
Reflectance Error
2-12 WARNING 1. Correctly ensure the reflectance value
and correction value with the blank
pad.
2. The subcategory codes represent the
items listed as follows:
Subcategory Item
2pH
3LEU
4NIT
5PRO
6GLU
7KET
8UBG
9BIL
10ERY
11COL
12CLA
35 C Setting?
Compensation Error
2-10 WARNING 1. Correctly ensure the correction value
with the blank pad.
2. The subcategory codes represent the
items listed as follows:
Subcategory Item
2pH
3LEU
4NIT
5PRO
6GLU
7KET
8UBG
9BIL
10ERY
36 Print Order
Print Error
1-12 WARNING 1. Print order setting is incorrect.
2. Correct the print order setting.
37 Control Position
Control Position Error
1-5 WARNING 1. Control level setting is incorrect.
2. Correct the position of the control
sample.
38 Parameter?
Parameter Error
1 WARNING 1. Contact Technical Support.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)