Roche Diagnostics
E-18 Operators Manual · Version 1.0
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
40 Sample Barcode
Reader
Sample Barcode
Reading Error
1STOP
WARNING
1. Check whether the barcode label is
dirty.
2. Check whether the barcode is
positioned correctly.
3. Check whether the barcode label is
missing.
4. Check whether the barcode label has
the correct type of code.
5. Contact Technical Support.
41 System I/F
System I/F Error
1-17
21-29
31-42
WARNING 1. Check the host connection and
communication.
2. Switch off the analyzer, then restart the
analyzer.
3. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
4. Contact Technical Support.
42 Touch Panel I/F
Touch Panel I/F Error
1-17 S.STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, contact
Technical Support.
21-29, 31-34 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, contact
Technical Support.
43 Memory Error
S Ram Error
1- 2 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
44 Initialize FLASH
Memory
Flash Memory Error
1 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)