EasyManua.ls Logo

Roche URISYS 2400 - Page 141

Roche URISYS 2400
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
Operators Manual · Version 1.0 E-19
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
45 Defragment Memory
Flash Memory Error
1 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
46 Data Area full
Flash Memory Error
1-3 S. STOP 1. Remaining capacity of the routine
sample data storage is 16 or less
(subcategory 1) at Start.
2. STAT sample data storage is six or less
(subcategory 2) at Start.
3. Control sample data storage is 10 or
less (subcategory 3) at Start.
4-6 WARNING 1. Remaining capacity of the routine
sample data storage is 45 or less
(subcategory 4) at Start.
2. STAT sample data storage is 27 or less
(subcategory 6) at Start.
3. Execute the data delete procedure after
backing up the old sample data to a
diskette.
47 FLASH Memory File
open
Flash Memory Error
1-9 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
48 FLASH Memory File
read
Flash Memory Error
1-9 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)

Table of Contents

Related product manuals