9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
49 FLASH Memory File
write
Flash Memory Error
1-9 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
50 FLASH Memory File
delete
Flash Memory Error
1-9 STOP 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
51 FLASH Memory
Flash Memory Error
1-10 STOP 1. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
2. Contact Technical Support.
52 FLASH Memory
Error
Flash Memory Error
1 WARNING 1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
53 No Floppy Disk
FDD Error
1 WARNING 1. Insert a diskette.
2. Format a diskette.
54 FD Write Protect
FDD Error
1 WARNING 1. Insert a diskette.
2. Release write-protection on a diskette.
55 FD Format
FDD Error
1 WARNING 1. Perform Clean FD Drive.
2. Perform Format FD.
56 FD Cleaning
FDD Error
1 WARNING 1. Perform Clean FDD Drive.
2. If the alarm persists, contact Technical
Support.
57 Floppy Disk read
FDD Error
1-4 WARNING 1. Perform Clean FD Drive.
2. If the alarm persists, contact Technical
Support.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)