URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
58 Floppy Disk write
FDD Error
1-5 WARNING 1. Perform Clean FD Drive.
2. If the alarm persists, contact Technical
Support.
59 Printer
Printer Error
1 WARNING 1. Check whether the printer is switched
on.
2. Check whether there is paper in the
printer.
3. Check the power supply and
communication cable to the printer.
60 TS Transfer Base
Test Strip
Transfer Base Error
1 WARNING 1. Check the position of the test strip
transfer base.
61 Rack Transfer
Rack Crossfeed Line
Error
1-3 STOP 1. Perform reset.
2. Use a different rack.
3. Contact Technical Support.
62 W.Syringe
Wash Syringe Error
1-2 STOP 1. Perform the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
63 Rack full
Rack Output Line Full
1 STOP 1. At Start/Reset:
2. Remove racks from the rack output tray
in the rack transfer mechanism, and
replace with an empty tray.
3. In Operation modes, other than Start/
Reset:
4. Same as before (63-1).
2STOP
4 WARNING 1. At Restart (from S.STOP):
2. Same as before (63-1/2)
5-6 STOP 1. Contact Technical Support.
7 WARNING 1. Place the rack tray in position
64 TS Set Error
Test Strip Cassette
1 WARNING 1. Execute the Change cassette
procedure again from the
[Maintenance] screen.
65 Wash Error
Wash Error
1 WARNING 1. Execute the wash procedure again.
66 Rack Tray
Rack Tray Error
WARNING 1. There is no rack tray on the output line.
2. Place an empty rack tray on the output
line of the rack transfer mechanism.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)