E-22 Operators Manual · Version 1.0
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
67 Sample Empty
No Sample Error
1WARNING1.No sample detected
68 Rack ID?
Rack ID Error
1 STOP 1. Check rack ID label.
71 Intensity?
Reflectance Error
1-7 STOP 1. Repeat calibration.
2. Contact Technical Support.
72 Factory Calibration?
Factory Cal Error
1-5 STOP 1. Contact Technical Support.
73 Calibration?
Calibration Error
1-2
3
STOP
WARNING
1. Repeat calibration.
2. Use new calibration strip.
3. If calibration is unsuccessful after
several attempts, contact Technical
Support.
74 Test Strip
Reflectance Error
1 WARNING 1. Check if urine is very dark. If not, repeat
measurement.
2. If alarm persists and urine samples are
not dark, contact Technical Support.
2-3 STOP 1. Repeat measurement.
2. Contact Technical Support.
78 SG?
Specific Gravity Error
1-6 WARNING 1. Carry out wash procedure.
2. Contact Technical Support.
79 SG Calibration?
Specific Gravity
Calibration 1 Error
1-6 STOP 1. Check the position of the SG standard.
2. Repeat SG calibration.
3. Carry out wash procedure.
80 SG Calibration
Check?
Specific Gravity
Calibration 2 Error
1-9 WARNING 1. Check the position of the SG standard.
2. Repeat SG calibration.
3. Carry out wash procedure.
81 No SG Calib Data
Specific Gravity
Calibration not done
1 WARNING 1. Execute SG calibration.
82 Specific Gravity
Calibration out of
range
SG out of range
1 WARNING 1. Carry out wash procedure.
2. Execute calibration.
3. Contact Technical Support.
83 SG?
Specific Gravity
System Error
1 WARNING 1. Execute calibration.
2. Contact Technical Support.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)