B-20 Operators Manual · Version 1.0
3 Software URISYS 2400
Installation
6 Press <Enter> after each entry.
7 Press <PgUp> to return to the [Controls] screen.
Mark the specified rack as a control rack. Do not use the control rack for
routine samples, otherwise the routine sample results will be stored in the
control data memory.
a To specify the control name and lot number
1 Press <Control Name/Lot No.> on the [Controls] screen. The [Control
Name/Lot No.] screen is displayed.
2 Press <Level> to select a level, for example, Level 1, Level 2 or Level 3.
3 Type the name of the control material, using a maximum of 16
alphanumeric characters. Press <Enter>.
4 Type the Lot No., using a maximum of 10 digits.
5 Press <Enter> after each entry.
6 Press <Level> again to enter the Name and Lot No. for the remaining
levels.
7 Repeat steps 3-5.
8 Press <PgUp> twice to return to the [Test Parameters] screen.