Network Connections and Remote Operation
R&S
®
RTB2000
157User Manual 1333.1611.02 ─ 04
2. Open the "Setup" menu.
3. Tap "Interface".
4. Tap "USB" to select USB connection.
5. Tap "Parameter".
6. Select the USB mode.
USB TMC (Test & Measurement Class)
USB VCP (Virtual Com Port)
USB MTP (Media Transfer Protocol)
USB TMC
USB TMC means USB Test & Measurement Class. You can use this interface for
remote control of the instrument using SCPI commands. USB TMC requires a VISA
installation on the controlling computer. VISA is used to access the instrument and to
send remote commands.
Rohde & Schwarz provides the standardized I/O software library R&S VISA for com-
munication via TCP/IP (LAN: HiSlip, VXI-11) or USB (USBTMC) interfaces. R&S VISA
is available for download at the Rohde & Schwarz website www.rohde-schwarz.com/
rsvisa.
USB VCP
USB VCP uses the virtual COM port (VCP) to communicate with the measuring instru-
ment. You can use any terminal program to send SCPI commands. USB VCP requires
the installation of the USB VCP driver on the controlling computer. The driver is availa-
ble for download at the Rohde & Schwarz website (www.rohde-schwarz.com/de/
treiber/hmo/, HO732).
USB MTP
USB MTP is the USB media transfer protocol. It is a solution to load data from the
oscilloscope to the computer. USB MTP does not need a driver installation. If the
instrument and the computer are connected with a USB cable and USB MTP is
selected as interface, you can see the oscilloscope in the computer's Device Manager
as portable instrument.
USB Connection