EasyManua.ls Logo

Rohde & Schwarz R&S RTB2004 - Vertical Settings

Rohde & Schwarz R&S RTB2004
509 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
RTB2000
252User Manual 1333.1611.02 ─ 04
Parameters:
<AcquisitionState> RUN | STOPping | COMPlete | BREak
RUN | STOP | COMPlete | BREak
RUN
Set: Starts the acquisition.
Read: The acquisition is running.
STOPping
Set: Stops the acquisition when it is finished.
Read: Acquisition is stopped.
COMPlete
Set: Not available.
Read: The current acquisition is finished and completed.
BREak
Set: Immediate interrupt of current acquisition.
Read: acquisition is finished but interrupted.
15.4.3 Vertical Settings
CHANnel<m>:STATe..................................................................................................... 252
CHANnel<m>:AON........................................................................................................ 253
CHANnel<m>:AOFF.......................................................................................................253
CHANnel<m>:SCALe..................................................................................................... 253
CHANnel<m>:RANGe.................................................................................................... 253
CHANnel<m>:OFFSet....................................................................................................254
CHANnel<m>:COUPling.................................................................................................254
CHANnel<m>:BANDwidth...............................................................................................254
CHANnel<m>:POLarity...................................................................................................255
CHANnel<m>:SKEW......................................................................................................255
CHANnel<m>:ZOFFset[:VALue]...................................................................................... 255
CHANnel<m>:OVERload................................................................................................ 256
CHANnel<m>:THReshold............................................................................................... 256
CHANnel<m>:THReshold:FINDlevel................................................................................256
CHANnel<m>:THReshold:HYSTeresis.............................................................................257
CHANnel<m>:LABel.......................................................................................................257
CHANnel<m>:LABel:STATe............................................................................................257
CHANnel<m>:STATe <State>
Switches the channel signal on or off.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<State> ON | OFF
Waveform Setup

Table of Contents

Related product manuals