Parameters:
<Width> Range: Depends on various settings, mainly time base and
sample rate
*RST: 400e-6
Default unit: s
SEARch:TRIGger:RUNT:DELTa <DeltaWidth>
Sets a range Δt to the reference pulse width set with SEARch:TRIGger:RUNT:WIDTh
, if SEARch:TRIGger:RUNT:RANGe is set to WITHin or OUTSide.
Parameters:
<DeltaWidth> Range: Depends on various settings, mainly time base and
sample rate
*RST: 100e-6
Default unit: s
15.6.4.7 Data2Clock Search Configuration
SEARch:TRIGger:DATatoclock:CSOurce......................................................................... 290
SEARch:TRIGger:DATatoclock:CLEVel............................................................................290
SEARch:TRIGger:DATatoclock:DLEVel............................................................................290
SEARch:TRIGger:DATatoclock:CLEVel:DELTa.................................................................291
SEARch:TRIGger:DATatoclock:DLEVel:DELTa.................................................................291
SEARch:TRIGger:DATatoclock:CEDGe........................................................................... 291
SEARch:TRIGger:DATatoclock:HTIMe.............................................................................291
SEARch:TRIGger:DATatoclock:STIMe.............................................................................291
SEARch:TRIGger:DATatoclock:CSOurce <ClockSource>
Selects the input channel of the clock signal.
Parameters:
<ClockSource> CH1 | CH2 | CH3 | CH4
*RST: CH1
SEARch:TRIGger:DATatoclock:CLEVel <ClockLevel>
Sets the voltage level for the clock signal. The crossing of clock level and clock edge
defines the start point for setup and hold time.
Parameters:
<ClockLevel> Range: Depends on vertical scale
SEARch:TRIGger:DATatoclock:DLEVel <DataLevel>
Sets the voltage level for the data signal. The data lavel defines the threshold for data
transition.
Waveform Analysis