453User Manual 1333.1611.02 ─ 04
Parameters:
<DutyCycle> Numeric value
Range: 1.00E+00 to 9.900E+01
*RST: 5.000E+01
Default unit: %
Example: PGEN:PATT:SQU:DCYC 20
Sets the duty cycle of the square wave function to 20%.
PGENerator:MANual:STATe<s> <State>
Selects the pins S0 to S3 manually and sets their states to high (H) or low (L).
Suffix:
<s>
.
0..3
Pins S0 to S3
Parameters:
<State> ON | OFF
ON
Pin state is set to high (H).
OFF
Pin state is set to low (L).
*RST: OFF
Example: PGEN:MAN:STAT2 ON
Sets the state of pin S2 to high (H).
15.14 Status Reporting
15.14.1 STATus:OPERation Register
The commands of the STATus:OPERation subsystem control the status reporting
structures of the STATus:OPERation register:
See also:
Chapter C.1, "Structure of a SCPI Status Register", on page 473
Chapter C.3.3, "STATus:OPERation Register", on page 478
The following commands are available:
STATus:OPERation:CONDition?..................................................................................... 454
STATus:OPERation:ENABle........................................................................................... 454
STATus:OPERation:NTRansition.....................................................................................454
STATus:OPERation:PTRansition.....................................................................................454
STATus:OPERation[:EVENt]?......................................................................................... 454