Remote Control Commands
R&S
®
RTB2000
287User Manual 1333.1611.02 ─ 04
SEARch:MEASure:LEVel:PEAK:MAGNitude <Magnitude>
Sets the peak-to-peak limit. If the signal exceeds this limit, a search event is listed.
Parameters:
<Magnitude> Default unit: V
15.6.4.5 Rise/Fall Time Search Configuration
SEARch:TRIGger:RISetime:SLOPe................................................................................. 287
SEARch:TRIGger:LEVel:RISetime:LOWer........................................................................287
SEARch:TRIGger:LEVel:RISetime:UPPer........................................................................ 287
SEARch:TRIGger:RISetime:RANGe................................................................................ 287
SEARch:TRIGger:RISetime:TIME....................................................................................288
SEARch:TRIGger:RISetime:DELTa................................................................................. 288
SEARch:TRIGger:RISetime:SLOPe <Polarity>
Sets the slope to be found.
Parameters:
<Polarity> POSitive | NEGative | EITHer
POSitive: to search for rise time.
NEGative: to search for fall time.
EITHer: to search for rise and fall time.
*RST: POS
SEARch:TRIGger:LEVel:RISetime:LOWer <LowerLevel>
Sets the lower voltage threshold. When the signal crosses this level, the rise time mea-
surement starts or stops depending on the selected slope.
Parameters:
<LowerLevel> *RST: 400 mV
Default unit: V
SEARch:TRIGger:LEVel:RISetime:UPPer <UpperLevel>
Sets the upper voltage threshold. When the signal crosses this level, the rise/fall time
measurement starts or stops depending on the selected slope.
Parameters:
<UpperLevel> *RST: 600 mV
Default unit: V
SEARch:TRIGger:RISetime:RANGe <Range>
Sets how the measured rise or fall time is compared with the given limit(s).
To set the rise/fall time, use SEARch:TRIGger:RISetime:TIME.