Remote Control Commands
R&S
®
RTB2000
258User Manual 1333.1611.02 ─ 04
15.4.4 Probes
PROBe<m>:SETup:ATTenuation:UNIT............................................................................ 258
PROBe<m>:SETup:ATTenuation:MANual........................................................................ 258
PROBe<m>:SETup:ATTenuation:UNIT <Unit>
Selects the unit that the probe can measure.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Unit> V | A
PROBe<m>:SETup:ATTenuation:MANual <ManualAttenuation>
Sets the attenuation or gain of the probe.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ManualAttenuation> Range: 0.001 to 10000
*RST: 1
15.4.5 Horizontal Settings
TIMebase:SCALe...........................................................................................................258
TIMebase:POSition........................................................................................................ 259
TIMebase:REFerence.....................................................................................................259
TIMebase:ACQTime.......................................................................................................259
TIMebase:RANGe..........................................................................................................259
TIMebase:DIVisions?..................................................................................................... 259
TIMebase:RATime?....................................................................................................... 260
TIMebase:SCALe <TimeScale>
Sets the horizontal scale for all channel and math waveforms.
Parameters:
<TimeScale> Range: 1e-9 to 50; lower limits are possible if zoom or
FFT is enabled.
Increment: 1e-9
*RST: 100e-6
Default unit: s/div