Remote Control Commands
R&S
®
RTB2000
369User Manual 1333.1611.02 ─ 04
SYSTem:COMMunicate:INTerface:USB:CLASs
Selects the USB mode.
●
USB TMC (Test & Measurement Class)
●
USB VCP (Virtual Com Port)
●
USB MTP (Media Transfer Protocol)
Parameters:
<USBClass> TMC | VCP | MTP
15.10.5 Trigger Out
TRIGger:OUT:MODE..................................................................................................... 369
TRIGger:OUT:PLENgth.................................................................................................. 369
TRIGger:OUT:POLarity...................................................................................................369
TRIGger:OUT:MODE <OutputMode>
Defines which signals are generated at the AUX OUT front connector.
Parameters:
<OutputMode> OFF | TRIGger | REFerence | MASK | GENerator
OFF
No output
TRIGger
Outputs a pulse when the instrument triggers.
Outputs a 10 MHz reference frequency.
MASK
Outputs a pulse when a mask is violated. This function is only
available if a mask is specified.
GENerator
Outputs the waveform specified using the function generator
(requires option R&S RTB-B6)
*RST:
OFF
TRIGger:OUT:PLENgth <PulseLength>
Defines the pulse width of the pulse at the AUX OUT front connector (at trigger event
or mask viaolation).
Parameters:
<PulseLength> *RST: 1E-6
TRIGger:OUT:POLarity <Polarity>
Defines the polarity of the pulse at the AUX OUT front connector (at trigger event or
mask viaolation).
General Instrument Setup