Remote Control Commands
R&S
®
RTB2000
337User Manual 1333.1611.02 ─ 04
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Example:
DVM2:SOUR CH2
DVM2:TYPE DCRMs
DVM2:RES?
<-- 7.089E-01
An RMS measurement is performed on measurement place 2,
on channel 2. The result is 708.9 mV.
Usage: Query only
DVM<m>:RESult[:ACTual]:STATus?
Returns the result value and the status of the result.
The status is the decimal representation of a 4-bit register value:
Bit 0 = 1: result is valid
Bit 1 = 1: no result available
Bit 2 = 1: clipping occurs
Bit 3 = 1: no period found
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Return values:
<CurrentValue> Measured value
<Status> Decimal status value
Example:
DVM:SOUR CH1
DVM:TYPE MEAN
DVM:RES:STAT?
<-- 4.968E-01,5
The result value of the mean measurement on channel 1 is
496.1 mV. The result status is 5 (decimal) = 0101 (binary). That
means, the result is valid (bit 0 = 1), and the signal is clipped by
the limits of the ADC range (bit 3 = 1).
Usage: Query only
15.8.5 Trigger Counter
TCOunter:ENABle..........................................................................................................338
TCOunter:RESult[:ACTual]:FREQuency?......................................................................... 338
TCOunter:RESult[:ACTual]:PERiod?................................................................................338